The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
References
Configurations
No configuration.
History
01 Mar 2025, 18:15
Type | Values Removed | Values Added |
---|---|---|
Summary | (en) The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device. |
31 Jan 2025, 17:15
Type | Values Removed | Values Added |
---|---|---|
Summary | (en) Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device. | |
CVSS |
v2 : v3 : |
v2 : unknown
v3 : 7.5 |
31 Jan 2025, 16:15
Type | Values Removed | Values Added |
---|---|---|
References |
|
|
Summary |
|
30 Jan 2025, 19:15
Type | Values Removed | Values Added |
---|---|---|
New CVE |
Information
Published : 2025-01-30 19:15
Updated : 2025-03-01 18:15
NVD link : CVE-2025-0626
Mitre link : CVE-2025-0626
CVE.ORG link : CVE-2025-0626
JSON object : View
Products Affected
No product.
CWE
CWE-912
Hidden Functionality