The product does not require unique and complex passwords to be created
during installation. Using Philips's default password could jeopardize
the PACS system if the password was hacked or leaked. An attacker could
gain access to the database impacting system availability and data
integrity.
References
Link | Resource |
---|---|
http://www.philips.com/productsecurity | Product |
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-200-01 | Third Party Advisory US Government Resource |
http://www.philips.com/productsecurity | Product |
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-200-01 | Third Party Advisory US Government Resource |
Configurations
History
09 Apr 2025, 21:16
Type | Values Removed | Values Added |
---|---|---|
CVSS |
v2 : v3 : |
v2 : unknown
v3 : 6.8 |
Summary | (en) The product does not require unique and complex passwords to be created during installation. Using Philips's default password could jeopardize the PACS system if the password was hacked or leaked. An attacker could gain access to the database impacting system availability and data integrity. |
Information
Published : 2024-07-18 17:15
Updated : 2025-04-09 21:16
NVD link : CVE-2023-40704
Mitre link : CVE-2023-40704
CVE.ORG link : CVE-2023-40704
JSON object : View
Products Affected
philips
- vue_pacs
CWE